Syringe Size |
10 ml, 20 ml, 30 ml, 50 ml/60 ml |
Applicable Syringe |
Compatible with syringe of any standard |
VTBI |
1 ml ~ 1000 ml (0.1, 1, 10 ml increments) |
Flow Rate |
5 ml: 0.1 ml/h ~ 100 ml/h (0.1, 1, 10 ml/h increments)
10 ml: 0.1 ml/h ~ 300 ml/h
20 ml: 0.1 ml/h ~ 600 ml/h
30 ml: 0.1 ml/h ~ 800 ml/h
50/60 ml: 0.1 ml/h ~ 1200 ml/h |
Bolus Rate |
5 ml: 100 ml/h ( in 0.1, 1, 10 ml/h increments)
10 ml: 300 ml/h
20 ml: 600 ml/h
30 ml: 800 ml/h
50/60 ml: 1200 ml/h |
Safety Classification |
Class II, Type BF |
Anti-Bolus |
Automatic |
Accuracy |
Volume accuracy : within ±2%
Mechanical accuracy : within ±1% |
Infusion Mode |
Flow rate: ml/min, ml/h
Time-based
Body weight: mg/kg/min, mg/kg/h, µg/kg/min, µg/kg/h etc. |
KVO Rate |
0.1-1 ml/h ( 0.01 ml/h increment ) |
Alarms |
Occlusion, near empty, end program
Low battery, end battery, AC power off
Motor malfunction, system malfunction
Reminder alarm, Pressure sensor error
Syringe installation error, Syringe drop off |
Drug Library |
Available |
Occlusion |
HIGH, MED, LOW |
Memory Log |
50000 events stored |
Power Supply, AC |
110 / 230 V (optional), 50-60 Hz, 20 VA |
Battery |
14.8V, Rechargeable |
Battery Life |
8 hours at 5 ml/h |
Working Temperature |
5? ~ 40? |
Relative Humidity |
20% - 90 % |
Atmospheric Pressure |
700 hPa ~ 1060 hPa |
Dimensions |
245 × 120 × 115 mm |
Weight |
3 kgs |