Syringe Size |
5 ml, 10 ml, 20 ml and 50 ml 60 ml Syringe |
VTBI |
0.1 ml ~ 9999 ml
0.1 ml ~ 999.9 ml increment: 0.1 ml
1000 ml ~ 9999 ml, increment: 1 ml |
Flow Rate |
5 ml 0.1 ml/h~150 ml/h
10 ml 0.1 ml/h~300 ml/h
20 ml 0.1 ml/h~600 ml/h
30 ml 0.1 ml/h~900 ml/h
50 ml 0.1 ml/h~1500 ml/h |
Bolus Rate |
if the injection rate is less than 800 ml/h, then bolus rate is 800 ml/h
if the injection rate is greater than or equal to 800 ml/h, the bolus rate is the maximal injection rate
5 ml : 150 ml/h
10 ml : 300 ml/h
20 ml : 600 ml/h
30 ml, 50 ml/60 ml : 800 ml/h |
Safety Classification |
Class I, Type CF |
Accuracy |
Volume accuracy : ±2 % or ± 0.005 ml/h |
KVO Rate |
0.1 ml/h ~ 5 ml/h |
Alarms |
Occlusion, completion, near empty, near completion, standby end, KVO completion
Syringe empty, Syringe-off, System failure, System abnormality, mis-operation, uninstalled syringe, communication failure,
Low battery, battery exhausted, power interruption |
Occlusion |
LOW: 40 kPa ± 16.7 kPa
MED: 70 kPa ± 16.7 kPa
HIGH: 120 kPa ± 24.0 kPa |
Time |
0-99 hour, 59 minutes |
Waterproof level |
IP34 |
Power Supply, AC |
AC: 100 V ~ 240 V (50/60 Hz,0.4-0.14A) DC10-16 V (2.25-1.5A) |
Battery |
Rechargeable battery, DC 7.4 V, 2600 mAh |
Battery Life |
6 hours at 5 ml/h after fully charged |
Working Temperature |
5? ~ 40? |